74 - Documenting Calibrator
Calibration results are saved and they can be viewed in MC6-T and also
sent to calibration software for further analysis.
Open
Documenting Calibrator
by tapping on the
Documenting
Calibrator
icon in MC6-T's Home View (see
Documenting Calibrator selected
).
Calibration Software
When this manual was printed, the following calibration software
supported communication with MC6-T:
•
Beamex CMX Calibration Software
, version 2.12 and later versions of
CMX Professional and Enterprise editions.
•
Beamex LOGiCAL
, version 2, a cloud based calibration management
software.
Note:
Older Beamex calibration software do not support MC6-T.
Calibrating Instruments
Instruments are typically calibrated following the procedure shown in
Figure 58: Typical calibration procedure
.
In MC6-T, you first select (or create) the instrument to be calibrated.
Then perform the As Found calibration – as many repeats as is required
- and decide whether or not adjustment is needed. Often you then do the
required amount of As Left repeats to document the instrument's state
after the possible adjustment.
A calibration procedure using MC6-T is presented in the following
subchapters.