Calibrating Instruments - 79
If RTD Temperature or TC Temperature is selected, the option Automatic
Control will appear. Calibrator will set the calibration points.
Continue the instrument setup by modifying several pages of instrument
configuration data. Note that by default only basic pages are shown. View
all pages by selecting
Show, All Pages
from the menu available when
creating/editing an instrument.
Basic pages
First page of the creating new instrument configuration data is
Input
related. For all input quantities (except Keyed option), parameters
Port /
Function
and
Unit
are available. Depending on selected
Quantity
and
Port / Function, other fields will become available for setup (see table
Table 1: Input parameters depending on the selected Quantity
Table 1: Input parameters depending on the selected Quantity
Quantity
Additional available settings
RTD Temperature
Sensor Type, Automatic control
TC Temperature
Sensor Type, RJ Mode, Automatic
control
Pressure
Pressure Type
Current
Loop Supply
Frequency
Amplitude, Waveform & Duty Cycle,
Trigger Level
Pulse
Amplitude, Waveform & Duty Cycle,
Frequency, Trigger Level, Trigger
Edge
Keyed
Keyed Quantity, Unit, Resolution,
Resolution from Entry