40 - Active Terminals and Connections
Note:
For HART and Fieldbus instrument specifics, see
. For information on External Devices (Pressure
and Temperature Controllers) used together with Calibrator and
Documenting Calibrator, see
.
In this manual, when presenting each function, there is a picture
with highlighted active terminals. The highlight for possible
optional connections is lighter. Connections to instruments are
included if it requires special attention. See chapter
.
Note:
A measurement, generation or simulation in one
subwindow reserve terminals in MC6-T. This may affect the
availability of measurements, generations and simulations in the
other subwindow. To free terminals, tap on the quantity button
and in the opened window, tap on the "
Stop
" button.
Measurements
Temperature Measurement (Thermocouple)
MC6-T has two thermocouple connectors.
TC1
is for cables and standard
TC plugs.
TC2
is for TC plugs with flat contacts.
Check the
Sensor Type
. Your measurement results are unreliable unless
you select the same sensor type as is connected to MC6-T. Select also
a suitable
Reference Junction
compensation method. Wrong settings
yield useless measurement results. For more information see chapter
.
Figure 22: Thermocouple measurement terminals. Range depends on
sensor type
.