108 - Data Logger
Figure 102: Data Log configuration window - 2nd page
A third page is available for measurement channels only: You may define
a trigger that invokes the Data Log. When triggering is configured, a
Figure 103: Data Log configuration window - 3rd
.
In this example, all four triggering methods are active. It is not necessary
to use all four. When more than one trigger method is active, they are
combined with logical
OR
function, i.e. as soon as one of them become
TRUE, Data Logging will start.
As shown in
Figure 103: Data Log configuration window - 3rd page
example, Data Logging will start as soon as any of the following become
true:
• The measured value is larger than 600
• The measured value is smaller than 45
• The rate of change of the measured value is larger than 20
• The rate of change of the measured value is less than -20
Summary of Contents for MC6-T
Page 2: ......
Page 38: ...38 General description ...
Page 66: ...66 Tools ...
Page 138: ...138 Communicator ...
Page 150: ...150 Service and transportation instructions ...
Page 168: ...168 Additional Information ...
Page 172: ...172 Index ...