124 - Communicator
Figure 119: Example of a variable list in a HART Transmitter
Unfortunately the structure of data and naming conventions in HART
instruments vary between different makes and models. Thus there is
no single path from selecting the instrument in MC6-T to locating the
parameter.
Note:
Go to
Menu, Show, Measurable Variables
for easy way
to find parameters.
Refer to your instrument's manual to find the blocks where the parameters
are located.
Once the parameter is located, tap on it. In the opened window: Tap the
Accept
button to start calibrating or logging that parameter.
Calibrating or Data Logging Fieldbus Instruments
This is a quick guide for selecting a
FOUNDATION Fieldbus
or
Profibus
PA instrument's Parameter
for calibration/data logging. If you want to
make more extensive configurations, refer to chapter
and the manual of your instrument at hand.
1.
First you must select either
FOUNDATION Fieldbus
or
Profibus
PA
as your communication protocol and select the instrument as
.
2. Transducer
is the typical name of the
Block
(folder) where
measurement parameters are located. When applicable, tap on
it. Note that the
Block
name depends on the instrument, so when
uncertain, refer to your instrument's user manual.