78 - Documenting Calibrator
Figure 64: Temperature Instrument Wizard, Input Method
Available methods for calibrating temperature instruments are:
•
Generated/Controlled
(available when Temperature Block is enabled)
- calibrator generates the temperature with internal Temperature Block
or controlls the temperature generation by an external device. This
method can be used with internal or external reference sensor (see
Figure 65: Documenting Calibrator, reference sensor selection
);
•
Simulated
- calibrator simulates the temperature;
•
Measured
- calibrator measures the temperature but does not control it;
•
Keyed
- measurement value is manually entered into the calibrator;
Figure 65: Documenting Calibrator, reference sensor selection dialog