130 - Communicator
Figure 126: Fieldbus instrument's trimming window, example
Refer to the instrument's user manual for detailed information on how the
trim procedure progresses. Usually you first set the
Mode Block
to OOS
(Out Of Service) and then start editing the other data.
The actual trim is done by tapping on
Trim 0 %
or
Trim 100 %
values after
a valid input signal is generated/simulated/measured. A window for editing
the numeric value opens. The window includes a button for copying the
input reading as the correct value.
Figure 127: Copy button
Note:
Some fieldbus instrument manuals and user interfaces
use the term "calibration" when they actually mean a trimming
procedure. Be aware of this pitfall when trimming a fieldbus
instrument.
Be careful during the trimming procedure. Follow the instructions
in the instrument's user manual. Skipping a step in the
procedure may result in an unsuccessful trim.
Summary of Contents for MC6-T