User Defined Function Data - 167
Figure 162: Example of a Controller Preset
The menu in a
Controller Preset
window contains further tools
for defining the communication. One of the menu items is
Enable
Measurement Port
, defining whether the is controller also used as a
measuring device. Menu selections may be individual for each preset.
When an External Controller is connected and configured, it is available as
any port in MC6-T's user interface.
Note:
If you run into communication problems, check the
communication settings in the connected controller. They may
have been changed from the controller's default values. MC6-T
always uses the controller's default values.
The Controller Preset window's menu includes an option to view
the
Communication Log
. In the Communication Log window
you may choose the log format: binary or text.
Changing Controller During Calibration
In Documenting Calibrator, you may need to change the controller during
calibration. This, e.g. because a temperature controller's span is not useful
throughout the measurement span of the instrument to be calibrated.
Change the controller as follows:
1.
During calibration, open the
Documenting Calibrator
's Menu.
2.
Select option
Instrument Input
.
3.
In the
Instrument Input
window, tap on the button that invokes the
controller change procedure.
4.
Follow the instruction on the screen.