5 Troubleshooting
5.1 ASSKEA secretion aspirators S-series
Error
Possible causes
Remedy
Device
does not
start.
Mains plug not plugged in
Check the electrical plug connection.
No mains voltage
Please contact Service! (fuse defective)
Device is too warm; temperature
protection of the motor has been
triggered.
Allow the device to cool down; temperature protec-
tion automatically resets after cooling.
insufficient
flow rate
External bacterial filter polluted,
moistened
Replace the external bacterial filter.
Integrated bacterial filter polluted,
moistened
Please contact Service!
Leak in the tubing system
Check that secretion canister lid and tubing connec-
tions are secure and no leakages are apparent.
Vacuum regulator (additional air
valve) not closed
Close the vacuum regulator (additional air valve)
(fig. 1(E)).
no flow
rate
Manual overflow protection of the
reusable secretion canister system
is closed.
Check the fluid level in the secretion canister; empty
the canister.
Hydrophilic overflow protection of
the disposable secretion canister
system is blocked.
Replace the disposable liner ‘OneWay’.
External bacterial filter blocked
Replace the external bacterial filter.
Integrated bacterial filter blocked
Please contact Service!
Suction tube blocked
Clean/replace the suction tube.
Contact ASSKEA GmbH or your ser-
vice partner if a malfunction cannot be
corrected by the described measures.