5.2 ASSKEA secretion aspirators M-series
Error
LED display
Possible causes
Remedy
Device does not
start.
LEDs do not
illuminate
Battery is discharged.
Connect the power supply unit for charging
or operating.
LED illuminates
red
Invalid input voltage
Connect the correct power supply unit; check the power
supply.
Device turns
itself off (applies
to M20).
LED blinks red,
beeper on after 10 s
Temperature of the battery too high
Switch off the device, if applicable, and allow it to cool
down. Connect the power supply unit and switch on the
device again.
LED switches off
Motor turns
itself off (applies
to M20D, M30D).
LED blinks red,
beeper on after
10 s, then the de-
vice turns itself off
after a further 20 s
Temperature of the battery too high
Switch off the device, if applicable, and allow it to cool
down. Connect the power supply unit and switch on the
device again.
LED switches off
Device only runs
with maximum
vacuum (applies to
M20D, M30D).
LED illuminates
red
Battery is discharged.
Switch off the device for charging after the aspiration
procedure.
LED illuminates
green when power
supply unit is
connected
Device does not
charge.
LED blinks red
Temperature of the battery too high for
charging
Allow the device to cool down. Automatic resumption of
charging when battery has cooled down.
LED illuminates
green
LED illuminates
red
Battery is defective.
Please contact Service!
LED illuminates
green
LED illuminates
red
Invalid input voltage
Connect the correct power supply unit,
check the power supply.