3.2.1 Connecting the reusable secretion canister system
1. Screw the lid of the reusable secretion canister sufficiently tight and attach the
canister to the device.
2. Insert the connecting tube with the blue elbow fitting
toward the side of the device. Make sure that the side of
the external bacterial filter with the lettering ‘IN’ is facing
the canister (directed toward patient)!
3. Place the grey elbow adapter of the connecting tube onto
the connection in the middle of the reusable secretion
canister lid (above the overflow protection).
4. Place the suction tube onto the second connection
(patient connection). Ensure that the tube is fixed
sufficiently tight.
Fig. 8
Fig. 9
Fig. 10
The reusable secretion canister or the holder
for the external canister ‘Bag’ is attached by
directly pushing it onto the guide rails on the
housing of the device (figs. 11, 12).
Please note that up to four mounting
positions can be used for the secretion
canister (figs. 11, 12) on all devices.
Fig. 11
Fig. 12