20
ENGLISH
3.2.2 Connecting the disposable secretion canister system
Malfunction due to collapsing disposable liner ‘OneWay’
A leak in the external canister ‘Bag‘ or at the lid of the disposable
liner ‘OneWay‘ resp. damage of the canister or the T-piece may
cause air to flow into the external canister ‘Bag’. This may lead
to the collapse of the disposable liner ‘OneWay’.
• Inspect the disposable secretion canister system to ensure
that the lid of the disposable liner ‘OneWay’ is securely
connected to the external canister ‘Bag’.
• Ensure that all connections are firmly attached and properly
connected.
• Ensure that the external canister ‘Bag’ is undamaged and the
T-piece is securely attached.
• Follow the instruction for use supplied by the manufacturer!
The original ASSKEA disposable secretion canister system consists
of the external canister ‘Bag’, the holder for the external canister
‘Bag’, the disposable liner ‘OneWay’, the connecting tube and the
suction tube.
Fig. 13
Connection designation:
A Vacuum connection
B Patient connection
Please also follow the
instruction for use sup-
plied with the dispo-
sable secretion canister
system!
Fig. 14
1. Remove the disposable liner
‘OneWay’ from the pack-
aging and fully extend it.