Instrument Setup 3-19
Upgrading the System Firmware
Introduction
You can update the GeneAmp PCR System 9700 firmware from a serial connection to
the RS485 port or from a PCMCIA Flash Memory Card.
Contact your local Applied Biosystems representative to obtain a PCMCIA Flash
Memory Card.
\
Upgrading Through
the Serial Port
The following procedure describes how to upgrade the firmware through the RS485
serial port.
To upgrade the firmware through the RS485 serial port:
Step
Action
1
Connect the appropriate communication cable from a computer to the RS485 port
on the 9700 instrument Base Module (see the figure below).
You can use the following cables:
For information on connecting the printer cable, see page 3-27.
2
From the
Diagnostics
screen, press
Upgrade
.
The Firmware Upgrade screen appears.
Communication cable
Part number
GR0877
RS485
Ports
Firmware Upgrade
PCcard - Upgrade from PC card
Serial - Upgrade from serial port
F1
F3
F4
F5
F2
Exit
Serial
PCcard