7-4 Routine Maintenance
Changing Fuses for
Double-Line
Operation
Instruments configured for double-line operation have two 8 amp Type T 250 V
5x20 mm fuses (
P/N 0999-1683).
To change the fuses for double-line instruments:
Step
Action
1
Turn off the instrument and disconnect the power cord from the side of the
instrument.
ELECTRICAL SHOCK HAZARD. Severe electrical shock, which
could cause physical injury or death, can result from working on an instrument when
the high voltage power supply is operating. To avoid electrical shock, disconnect the
power supply to the instrument, unplug the power cord, and wait at least 1 minute
before working on the instrument.
The fuses are located in the Power Entry Module.
2
Insert a small flat-tip screwdriver into the slot in the upper portion of the power entry
module, and open the door.
3
Pull the fuse compartment out.
There are two fuses in the fuse compartment.
4
Pull out the fuse from the back of the fuse compartment and replace the blown fuse
with one 8 amp Type T 250 V fuse.
5
Place the fuse compartment back into the Power Entry Module and close the door.
Press the door until it locks in place.
6
Connect the instrument power cord.
DANGER
!