Instrument Setup 3-5
Converting from a Single to Double-Line Fuse
Introduction
FIRE HAZARD. For continued protection against the risk of fire, replace fuses
only with listed and certified fuses of the same type and rating as those currently in the
instrument.
This procedure describes how to convert the instrument from the single-line
configuration to double-line service (220/230/240 V nominals only).
IMPORTANT
For proper operation, you must know the power source(s) available and
determine if the instrument fuse configuration is correct. The instrument is shipped configured
for single-line operation.
Fuse Part Number
The correct fuse is an 8 amp Type T 250 V 5x20 mm fuse, P/N 0999-1683
.
Power Entry Module
Diagram
The following is a diagram of the Power Entry Module.
WARNING
!
GR0961
Jumper clip
Fuse compartment
Compartment door
Fuse
GR0878
One fuse
already installed