Troubleshooting 8-3
Display Screen Error Messages
Error Messages
Table
Refer to the following table for a description of error messages, and recommended
actions that you should take.
Table 8-1
Error Messages
Message
Description
Recommended Action
Battery RAM version number
lost
This error is generated when the battery RAM has
been lost and re-initialized.
Call Technical Support.
Block Calibration initialized
Software or hardware failure.
Call Technical Support.
Block Calibration reset to
default
System error.
Block data reset to defaults.
Call Technical Support.
Block isn’t configured
The instrument has defaulted to a 96-well
configuration.
Call Technical Support.
Block initialized
Block module has been re-initialized.
Call Technical Support.
Block version unknown,
update firmware
The calibration data in the block is not recognized by
the firmware.
Upgrade the firmware.
Bus Error
System error.
Call Technical Support.
Calibration battery RAM
initialized
Calibration lost. Instrument may not perform to
specification.
Call Technical Support.
Can only enter an infinity hold
at end
A method can only have a HOLD segment with an
infinity hold as the last segment in a method.
This message occurs when you try to enter an infinity
hold segment in the middle of a method.
Assign finite time segments
to holds within a method.
Can’t allocate timer
System error.
Call Technical Support.
Delete your methods first
User tried to delete a user name that has methods
stored under it.
Delete or transfer the
associated methods before
deleting a user name.
Enter a name or CANCEL
You did not enter at least one character on the User
Name screen before pressing the Accept key.
Enter the user name to
which the desired method is
assigned.
Enter oligo sequence
Incomplete TmCalc data.
Enter a value in the P1P2
fields of the TmCalc.
Enter user and method names
or CANCEL
You did not enter a user name and a method name
before storing a method.
Specify the method name
and choose a user to store a
method.
FATAL – Block shut off by
hardware
Block thermal runaway.
Call Technical Support.
FATAL – Block thermal runway
Fatal error.
Turn off system.
Call Technical Support.
FATAL – Cover shut off by
hardware
Heated cover thermal runaway.
Call Technical Support.
FATAL – Heat sink is too hot
Ambient conditions may be too warm.
Call Technical Support.
FATAL – Heat sink sensor
failure
System error.
Call Technical Support.
FATAL – Heated cover thermal
runaway
Fatal error.
Turn off system.
Call Technical Support.
Summary of Contents for GeneAmp PCR System 9700
Page 1: ...GeneAmp PCR System 9700 Base Module User s Manual ...
Page 9: ...ix To Obtain Customer Training Information C 7 D Limited Warranty Statement Index ...
Page 10: ......
Page 28: ......
Page 56: ......
Page 90: ......
Page 96: ......
Page 110: ......
Page 116: ......
Page 124: ......
Page 126: ......
Page 131: ......