8-4 Troubleshooting
FATAL – Sample block sensor
failure
Fatal error.
Turn off system.
Call Technical Support.
Fatal – Stack Overflow
A warning or error message that displays which task
stack overflowed.
The warning message is issued when the stack has
reached within 10% of overflowing.
Call Technical Support.
Field is full
You tried to enter more data in a field than the field size
allows.
Reenter data within the
specifications of the field.
Heated cover sensor failure
The heated cover sensor failed.
Call Technical Support.
Infinity hold not allowed in
cycle
A method can have an infinity hold segment as the last
segment in the method.
This message occurs when you tried to enter an infinity
time in a CYCLE segment.
Use finite values for cycle
segments within the
method.
Invalid password/pin#
You entered an incorrect PIN#.
Enter the correct PIN#.
LCD screen timed-out
System error.
Display screen and firmware have a faulty connection.
Call a Technical Support.
List of user names is full
The maximum number of users has been entered into
the system.
Delete unused user names.
Maximum of 6 segments
allowed
You tried to insert more than six temperature control
parameters into a hold or cycle.
Do not assign more than six
hold or cycle parameters to
a method.
Method battery RAM initialized
Stored methods have been reset due to hardware or
software failure.
Not all methods may be lost.
Check method directory.
Call Technical Support.
Method requires at least one
segment
You deleted all temperature control parameters in a
method.
A method must have at least one time and temperature
parameter.
Review and correct the
method to include the
temperature parameter(s).
No seconds in time field
You did not include seconds in the time field.
Include seconds when
entering the time.
Not enough method memory
left
This error occurs:
♦
When you attempt to exceed the limit of 137
methods.
♦
When you attempt to store or create a new method
which is larger than the available storage space.
♦
Determine how much
storage memory is
available on the
instrument or PC card.
♦
Delete or store rarely
used methods
elsewhere.
Not implemented yet
The feature is not implemented in the current firmware
version.
Upgrade firmware when the
new version is available
from Applied Biosystems.
PC card and Flash do not
verify
The PC (upgrade) card and instrument memory do not
match.
Firmware upgrade unsuccessful.
Call Technical Support.
PC card does not contain valid
data
The PC card being used to upgrade the instrument
does not contain a valid program.
Call Technical Support.
Table 8-1
Error Messages
(continued)
Message
Description
Recommended Action
Summary of Contents for GeneAmp PCR System 9700
Page 1: ...GeneAmp PCR System 9700 Base Module User s Manual ...
Page 9: ...ix To Obtain Customer Training Information C 7 D Limited Warranty Statement Index ...
Page 10: ......
Page 28: ......
Page 56: ......
Page 90: ......
Page 96: ......
Page 110: ......
Page 116: ......
Page 124: ......
Page 126: ......
Page 131: ......