Scanning Plates
CytoFluor User’s Guide
3-25
3
Running
1.
Click the Start button to start the run. If you start a run
before the lamp has been on for 10 minutes, you see a
screen message asking if you want to start the run.
Select Yes to continue or No to wait. If you select Yes,
you may not receive optimal results since the lamp was
not warmed up completely.
If you entered a file name or opened a file, you see a
screen message asking if you want to overwrite the
data.
Select Yes if the screen contains no data. Select No if
you do not want to overwrite the data. Then select New
from File and enter a new name using Save As….
As the system scans each well, you can see the AFU
value appear in the corresponding well at the bottom of
the screen.
Accessing other
applications
You can access other Microsoft
®
Windows
®
applications
while performing a scan.
To access other applications:
1.
Wait until the scan is between cycles.
2.
Double-click the icon of the application.
You can also hold down the ALT key and press Tab
to toggle among open applications.
Stopping a run
You can stop the run at any time by pressing the Escape key
on your keyboard. If you cancel the scan before it finishes, you
see:
Figure 3-9 Cancel Run