Overview of the Plate Layout Screen
CytoFluor User’s Guide
5-3
5
The Plate Layout screen includes the following features:
•
Well Types Options and Index Selection box—
Enable you to specify characteristics of particular wells
and define replicates. The system may use the data of
specific wells to perform calculations, for example,
background subtraction. See Section 5.4, Using Well
Types Options, for details.
The index selection box (located to the right of the Well
types) enables you to select an index value for replicate
wells (00-99) of the same type. The index value
enables you to identify replicates of the same material
on the plate. (You do not need to enter an index value
for wells defined as Empty, Hidden, or Excluded.)
•
Concentration parameters—Enable you to enter
concentration and dilution factors for each group of
wells.
•
Current Plate Description—Allows you to view the
size and coordinates of the plate you are using.
•
Basic editing options—Enable you to cancel your
entries or confirm them. See “Basic Editing Guidelines”
on page 5-6, for details.
•
Plate Representation (rows A to H columns 1 to 12
and well blocks)—Allows you to select the wells you
want to read. You can also view the results of your
selections as you make them.