Chapter 6 Customizing the CytoFluor System
6-24
Applied Biosystems
6
Figure 6-19 Checking Probe Height, Cutaway View
Adjusting probe
height in top-read
To adjust the probe height in the top-read position:
1.
Loosen the two screws on the rear of the probe
mounting block. Do not remove them. Leave the probe
connected to the mounting bracket.
2.
Slide the probe up or down to the correct height.
CAUTION
Bring the probe head down to just touch the 0.020-inch
spacer. Do not place excess force on the plate carrier,
or you could damage it.
Figure 6-20 Adjusting Probe Height
3.
Tighten the two screws on the rear of the probe mounting
block. Gently remove the 0.020-inch spacer.
Temperature
control chamber
0.020-inch spacer