Configuring the Scanner for Top or Bottom Plate Reading
CytoFluor User’s Guide
6-21
6
CAUTION
If you under-tighten or over-tighten the cover thumbwheel,
the plate carrier cannot exit the scanner. To make sure you
shut the cover correctly, click on the Plate Out button from
the Maintenance screen.
6.5.3 Checking and Adjusting Probe Height
in a Temperature Control Scanner
Use the following procedures to configure a scanner with a
temperature control chamber.
6.5.3.1 Checking and Adjusting Probe Height in
Bottom-Read Position (TC Models)
Checking probe
height in bottom-
read
In the bottom-read position, the probe tip should be located
0.4 mm (the thickness of the 0.015-inch height adjustment
spacer) below the plate carrier. You should rarely have to
reset the bottom-read probe height, because the depth of
different plates does not vary widely. If you need to adjust the
probe height, call Applied Biosystems Technical Support for
assistance.
CAUTION
The probe must be set at the correct height. If you set the
probe too high, the plate carrier can jam against it. If you
set it too low, you lower the signal.
1.
Turn off the scanner. Open the scanner cover by
turning the thumbwheel clockwise and lifting up the
cover.