Chapter 3 Scanning Operations
3-22
Applied Biosystems
3
3.5 Controlling the Temperature
CytoFluor Series
4000/TC
This section describes features unique to CytoFluor
Series 4000 systems with temperature control. If your system
does not include a temperature control chamber, skip this
section.
Overview
The CytoFluor version 3.1 and later software allows you to
set and read the temperature of CytoFluor Series 4000
models equipped with a temperature control chamber. You
can control the temperature to hold the plate at a specific
temperature or to perform kinetic experiments that measure
rates or endpoints.
Temperature
stability
The temperature you set has an accuracy of ± 0.5°C and is
uniform to ± 0.3°C.
System
temperature
The Temp box (Figure 3-7) displays the actual system
temperature and is updated every 5 seconds as the
temperature equilibrates. The temperature can take
15 to 20 minutes to stabilize. When scans are running, the
temperature is updated at the beginning of each scan.
Figure 3-7 Setting the Temperature
pull-down
SetTemp
Current
temperature
menu