Chapter 7 Maintaining the CytoFluor System
7-4
Applied Biosystems
7
7.3 Replacing the Air Filter
When to replace
Replace the air filter at least every six months. If your
installation site is dusty, you may need to change this filter
more frequently.
Replacing
1.
Turn off the CytoFluor scanner. Then remove the three
screws that secure the air filter protector on the rear
panel. Use the Allen (hex) wrench provided with the
Accessories kit.
Figure 7-2 Removing the Air Filter
2.
Remove and discard the used filter.
3.
Locate the replacement filter included with the kit. The
filter is in two 31 mm x 82 mm pieces; or one 31 mm x
164 mm piece. Insert the new filter and replace the air
filter protector. Tighten the screws with the Allen (hex)
wrench, but do not overtighten.
CAUTION
Do not operate the scanner without an air filter or you may
reduce the performance quality.
PB100543