CytoFluor User’s Guide
B-1
Error Messages
B
This appendix lists error messages you may see while using
the CytoFluor software. The CytoFluor software displays a
system error in a message box and makes a beep sound to
draw your attention to it. Many error messages display a
negative error code (number). This chart lists the possible
error codes or messages you may see and defines their
meaning:
Error Code/ Message
Meaning
-202
Indicates a bug in the program.
-203
Failure to initialize the PC’S Com port. The port
may not exist, or may already be claimed by
Windows or by another program.
-204
Time-out waiting for response from instrument. The
instrument may not be turned on or connected to
the PC, or the Com port may be incorrectly
configured. (See error code -206.)
-205
Communication aborted by ESC key. You pressed
the Escape key.