Chapter 6 Customizing the CytoFluor System
6-20
Applied Biosystems
6
Adjusting probe
height in top-read
1.
Loosen the two screws on the rear of the probe
mounting block. Do not remove them. Leave the probe
connected to the mounting bracket.
2.
Slide the probe up or down to the correct height.
CAUTION
Bring the probe head down to just touch the 0.020-inch
spacer. Do not place excess force on the plate carrier,
or you could damage it.
Figure 6-18 Adjusting Probe Height
3.
Tighten the two screws on the rear of the probe
mounting block. Gently remove the 0.020-inch spacer.
4.
Slide the plate carrier to the back of the scanner.
5.
Make sure the fiber-optic cables are inside the scanner.
CAUTION
The fiber-optic cables can be damaged if they are
caught between the cover and the base of the scanner.
6.
Close the scanner cover and seat it properly on the
gasket. Turn the thumbwheel counterclockwise until finger
tight. Turn on the scanner.
0.020-inch spacer
Fiber-optic cables