Replacing Optical Filters
CytoFluor User’s Guide
6-5
6
The scanner filter wheel finds the home position, then
presents the selected location at the top of the filter
assembly.
6.
Fold down the gasket in the filter assembly to gain access
to the filter (Figure 6-5). Then locate the filter installation
tool provided with the Accessories kit.
Figure 6-5 Accessing the Filter
7.
Place the prongs of the tool into the holes on the top of the
filter wheel or blank you need to change. Turn the tool to
the left until you can lift the filter from the wheel.
Figure 6-6 Removing the Filter
CAUTION
Do not apply excessive pressure while turning the filter;
you could damage the filter wheel. And do not touch the
glass surfaces of any filter or the instrument’s optical
components, or you could damage them.