Setting Auto Scale Parameters
CytoFluor User’s Guide
5-25
5
5.8 Setting Auto Scale Parameters
Overview
The CytoFluor Auto Scale screen allows you to select how you
normalize your data using the following modes:
•
Default mode—Scales data according to the filter
pair you set and the internal reference and gain the
scanner automatically sets.
•
Defined mode—Scales data according to the filter
pair settings, reference, and gain you set. The
defined mode allows you to set up to four different
combinations of Auto Scale parameters for each one of
four scans.
•
Auto Scale off—Collects data without auto scaling.
AutoScale off is the default setting.
NOTE: For the system to apply the settings, you must
select the Auto Scale field in the CytoFluor main setup
screen before you start a run. See “Auto Scale” on
page 3-13.