Chapter 7 Maintaining the CytoFluor System
7-6
Applied Biosystems
7
4.
Remove the lamp by pulling straight up on the lever to the
left of the lamp until it is free. Discard the faulty lamp.
Figure 7-4 Removing the Lamp
5.
Remove the new lamp from the box by grasping the
outside rim of the reflector.
CAUTION
Do not touch other areas of the new lamp or the interior
of the reflector. They must be free of fingerprints or
other material that can interfere with illumination and
collect dust. (You may want to wear lab gloves.)
Locate the tab on the lamp rim. Insert the new lamp into
the socket so that the it faces the excitation filter wheel
with the tab facing up. Press lightly on the rear of the
lamp to seat and align it.
Figure 7-5 Replacing the Lamp
6.
Place the cover over the lamp compartment. Then insert
and tighten the screw.