The CytoFluor Scanner
CytoFluor User’s Guide
2-7
2
Letter Part
Name
Function
A
Plate carrier
Holds the plate so the scanner can read the well
contents.
B Emission
filter
assembly
Holds the filter that determines the fluorescence
emission wavelength to monitor.
C
Probe
Transmits and receives the excitation and emission
signal.
D
Top read mounting
block
Holds the probe so it can read from the top of the
plate.
E
Bottom read
mounting block
Holds the probe so it can read from the bottom of
the plate.
F
Excitation filter
assembly
Holds the filter that selects the excitation
wavelength.
G
Lamp
Provides the fluorescent excitation. (The lamp is
located at the rear of the scanner in a separate,
covered compartment.)