Section 5
Remote Control
5-22
SYStem:COMMunicate:GPIB:HEAD
Function
Sets the addition of header.
Program message
SYSTem:COMMunicate:GPIB:HEAD < flag >
SYSTem:COMMunicate:GPIB:HEAD?
Response message
SYSTEM:COMMUNICATE:GPIB:HEAD < status >
Parameter
< flag >
:= {ON, OFF, 1, 0}
< status >
:= {1, 0}
1 ........... ON
0 ........... OFF
Explanation
Sets whether the header is appended at the response message, or not.
For the default value, the header is not appended.
There are the same setting items at GPIB and Serial (namely, System:
COMMunicate:GPIB:HEAD and System:COMMunicate:SERial:HEAD), and these
are dependent on each another. When the one is set, the another is set to the same.
SYStem:COMMunicate:SERial:HEAD
Function
Sets the addition of header.
Program message
SYSTem:COMMunicate:SERial:HEAD < flag >
SYSTem:COMMunicate:SERial:HEAD?
Response message
SYSTEM:COMMUNICATE:SERIAL:HEAD < status >
Parameter
< flag >
:= {ON, OFF, 1, 0}
< status >
:= {1, 0}
1 ........... ON
0 ........... OFF
Explanation
Sets whether the header is appended at the response message, or not.
For the default value, the header is not appended.
There are the same setting items at GPIB and Serial (namely, System:
COMMunicate:GPIB:HEAD and System:COMMunicate:SERial:HEAD), and these
are dependent on each another. When the one is set, the another is set to the same.
Summary of Contents for MT9812B
Page 11: ...For Safety xi Laser Radiation Markings D E C A B G F 1...
Page 20: ...xx...
Page 24: ...IV...
Page 38: ......
Page 52: ...Section 3 Before Use 3 14...
Page 68: ...Section 4 Operation 4 16...
Page 118: ...Section 5 Remote Control 5 50...
Page 136: ...Appendix App ii...
Page 142: ...Appendix A Performance Test Result Recording List A 6...
Page 144: ...Appendix B Error Code B 2...