Section 5
Remote Control
5-18
Table 5-4
Combinations of Measuring Range and Bandwidth for MU931311A
Range
0.1 Hz
1 Hz
10 Hz
100 Hz
1 kHz
10 kHz
100 kHz
+10 to
0 dBm
0
to
–10 dBm
–10
to
–20 dBm
–20
to
–30 dBm
–30
to –40 dBm
–40
to
–50 dBm
–50
to
–60 dBm
×
–60
to
–70 dBm
×
×
–70
to
–80 dBm
×
×
×
–80
to
–90 dBm
×
×
×
×
×
–90
to –110 dBm
×
×
×
×
×
HINT
It is recommended that ordinarily the bandwidth set-
ting be selected as “AUTO.” When using the MT9812B
with an arbitrary specified fixed bandwidth, the mea-
surement value can contain some error.
SENSe<n>:FILTer:BPASs:FREQuency
Function
Sets the modulation frequency.
Program message
SENSe<n>:FILTer:BPASs:FREQuency CW | < freq > [< unit >]
SENSe<n>:FILTer:BPASs:FREQuency?
Response message
< freq >
Parameter
< freq >
:= {0, 270, 1000, 2000} (Unit: Hz)
< unit >
:= {HZ, KHZ}
Explanation
This command sets the modulation frequency of the incident light to be measured.
When the units of the <freq> are omitted, they are set in Hz unit. When the units
are specified with the value of the <unit>, it is set in the specified one.
For 0 Hz, it is set as CW.
In the Program message, the value in either Hz or kHz can be set. But the re-
sponse message value is output only in Hz unit.
Summary of Contents for MT9812B
Page 11: ...For Safety xi Laser Radiation Markings D E C A B G F 1...
Page 20: ...xx...
Page 24: ...IV...
Page 38: ......
Page 52: ...Section 3 Before Use 3 14...
Page 68: ...Section 4 Operation 4 16...
Page 118: ...Section 5 Remote Control 5 50...
Page 136: ...Appendix App ii...
Page 142: ...Appendix A Performance Test Result Recording List A 6...
Page 144: ...Appendix B Error Code B 2...