5-3
Remote Control
5
5.1
Setttings
(2)
Using RS-232C
Pin 8: Sets to 0 for RS-232C.
Pin 7: Specifies the bit length. 0: 8 bits, 1: 7 bits
Pin 6: Specifies whether parity is valid or not. 0: Invalid, 1: Valid
When 1 (Valid) is specified, the pin-5 setting becomes valid.
Pin 5: Specifies the parity. 0: Odd, 1: Even
Pin 4: Sets the stop bit. 0: 1 bit, 1: 2 bits
Pin 3: Specifies the speed for S3.
Pin 2: Specifies the speed for S2.
Pin 1: Specifies the speed for S1.
Next table shows the speed setting
1200
2400
4800
9600
14400
19200
19200
19200
Speed (bps)
S3
0
0
0
0
1
1
1
1
S2
0
0
1
1
0
0
1
1
S1
0
1
0
1
0
1
0
1
Fig. 5-3 shows a RS-232C setting example.
GPIB/RS-232C
1
0
1
2
3
4
5
6
7
8
Fig. 5-3
RS-232C Setting Example
RS-232C, 8 bits, no parity, stop bit: 2 bits, 9600 bps
POINT
The changed DIP switch setting becomes valid after
the next power on. So, for using the MT9812B with the
changed switch setting, turn the power on again.