5-19
Remote Control
5
SENSe<n>:POWer:INTerval
Function
Sets the measuring interval.
Program message
SENSe<n>:POWer:INTerval < time >
SENSe<n>:POWer:INTerval?
Response message
SENSE<n>:POWER:INTERVAL < time >
Parameter
< time >
:= {f|0:001
≤
f
≤
359999}
Explanation
This command sets the measuring interval.
The interval can be set as 0.001 s (1 ms), 0.01 s (10 ms), 0.02 s (20 ms), 0.05 s (50
ms), 0.1 s (100 ms), 0.2 s (200 ms), 0.5 s (500 ms), or a value between 1 s and
359,999 s with 1 s resolution.
The <time> is set in units of seconds and rounded off by the resolution.
POINT
The MT9812B does not support the data read-out func-
tion for a measuring interval of 100 ms or shorter. So,
the measuring interval between 1 ms and 50 ms for the
optical sensor unit can be set, but the measured value
may not be read-out, correctly. When using the optical
sensor unit in the MT9812B, set the measuring interval
to 100 ms or longer.
SENSe<n>:MEMory:COPY[:NAME]
Function
Initializes the measuring condition of optical sensor unit.
Program message
SENSe<n>:MEMory:COPY[:NAME] 0, MC
Explanation
This command returns the measuring condition of the optical sensor unit to the
status at the shipment.
POINT
The MT9812B is not provided with the measuring-con-
dition saving function, but units are initialized by the
same message (reading out the contents in memory
“0”) as that used in the Optical Test Set MT9810A.
5.3
Device Message
Summary of Contents for MT9812B
Page 11: ...For Safety xi Laser Radiation Markings D E C A B G F 1...
Page 20: ...xx...
Page 24: ...IV...
Page 38: ......
Page 52: ...Section 3 Before Use 3 14...
Page 68: ...Section 4 Operation 4 16...
Page 118: ...Section 5 Remote Control 5 50...
Page 136: ...Appendix App ii...
Page 142: ...Appendix A Performance Test Result Recording List A 6...
Page 144: ...Appendix B Error Code B 2...