5-47
Remote Control
5
5.10.3 QUESTIONABLE Status Register
QUESTIONABLE status register has the following information.
• Output power of light source
• Abnormalities of current/temperature/power-supply
The QUESTIONABLE status register has the following commands, which are
inserted into the <node> field of the status register for use.
Command
Description
STATus:QUESTIONABLE:POWER
QUESTIONABLE status
register
STATus:QUESTIONABLE:POWER:SOURCE Output power of light source
STATus:QUESTIONABLE:CURRENT
Current abnormality
STATus:QUESTIONABLE:ENVTEMP
Temperature abnormality
STATus:QUESTIONABLE:POWER
Power supply abnormality
STATus:QUEStionable:POWer
Function
QUSTIONABLE status register
Explanation
Specifies the status register to be referenced.
Each device state is indicated at a bit allocated to it.
Each bit content is as follows.
Bit
Contents
2
Remote interlock
3
Output power of light source
6
Current abnormality
7
Temperature abnormality
8
Power supply abnormality
These bits have the low-order structure, excluding the remote interlock. When
the remote interlock is set to 1, the system is in the remote-interlock on state.
When it is set to 0, the system is in the remote-interlock off state.
STATus:QUEStionable:POWer:SOURce
Function
Indicates output ON/OFF state of light source.
5.10
Status Commands
Summary of Contents for MT9812B
Page 11: ...For Safety xi Laser Radiation Markings D E C A B G F 1...
Page 20: ...xx...
Page 24: ...IV...
Page 38: ......
Page 52: ...Section 3 Before Use 3 14...
Page 68: ...Section 4 Operation 4 16...
Page 118: ...Section 5 Remote Control 5 50...
Page 136: ...Appendix App ii...
Page 142: ...Appendix A Performance Test Result Recording List A 6...
Page 144: ...Appendix B Error Code B 2...