6-3
Performance Test
6
6.1
Light Source Performance Test
6.1.1
Optical Output Level
Measuring instrument tested
Optical power meter
Optical fiber
Fig. 6-1
Step
Operation
1
Set up a measurement system like that shown in the figure, above.
2
Set the optical output mode of the measuring instrument to be
tested to the CW mode, and set ATT at 0 dB.
3
Set the measuring instrument to be tested to the center frequency.
4
Measure the optical output level by the optical power meter.
5
Check that the value measured comes within the specified level.
6.1.2
Center Optical Frequency
Measuring instrument tested
Variable optical attenuator
Optical Frequency Counter
MF9630A
Optical fiber
Optical fiber
Fig. 6-2
Step
Operation
1
Set up a measurement system like that shown in the figure,
above.
2
Set the optical output mode of the measuring instrument to be
tested to the CW mode, and set ATT at 0 dB.
3
Set the measuring instrument to be tested to the center fre-
quency.
4
Adjust the optical output to or below the maximum input level
of the optical frequency counter using the optical attenuator.
5
Measure the optical frequency by the optical frequency counter.
6
Check that the value measured comes within the specified
value.
Summary of Contents for MT9812B
Page 11: ...For Safety xi Laser Radiation Markings D E C A B G F 1...
Page 20: ...xx...
Page 24: ...IV...
Page 38: ......
Page 52: ...Section 3 Before Use 3 14...
Page 68: ...Section 4 Operation 4 16...
Page 118: ...Section 5 Remote Control 5 50...
Page 136: ...Appendix App ii...
Page 142: ...Appendix A Performance Test Result Recording List A 6...
Page 144: ...Appendix B Error Code B 2...