5-15
Remote Control
5
FETCh<n>[:SCALar]:POWer[:DC]
Function
Inquiries the measured data.
Program message
FETCh<n>[:SCALar]:POWer[:DC]?
Response message
< level >
Parameter
< level >
:=< NR3 >
Explanation
This command returns the latest data measured at the time of issuing the message.
The response message is output in the unit value that is set in “SENSe<n>:
POWer:UNIT”
SENSe<n>:POWer:UNIT
Function
Switches the readout-data unit system.
Program message
SENSe<n>:POWer:UNIT < unit >
SENSe<n>:POWer:UNIT?
Response message
< unit >
Parameter
< unit >
:= {DBM, W}
Explanation
Switches the unit of the response to “FETCH<n>[:SCALar]:POWer[:DC]?.”
POINT
This command switches only the units of the response
message. The unit display on the front panel of the
MT9812B remains “dBm.” In addition, the unit of the
response message to “MFRame:FETCh:POWer?
<ch-list>” also remains “dBm.”