5-11
Remote Control
5
5.3.3
Commands for Setting Optical Sensor Unit
Using the optical sensor unit setting commands, the setting is made for the optical
sensor unit specified by <n>.
SENSe<n>:CORRection:COLLect:ZERO
Function
Executes offset.
Program message
SENSe<n>:CORRection:COLLect:ZERO
SENSe<n>:CORRection:COLLect:ZERO?
Response message
< result >
Parameter
< result >
:=< NR1 >
Explanation
This command executes offset processing. The response message has one of the
following values. See “SYSTem:ERRor” for information on error codes.
< result >
Status
0
Normal termination
1
Offset processing not executed by remote control.
2
Offset processing in progress
Negative number
Error
5.3
Device Message
Summary of Contents for MT9812B
Page 11: ...For Safety xi Laser Radiation Markings D E C A B G F 1...
Page 20: ...xx...
Page 24: ...IV...
Page 38: ......
Page 52: ...Section 3 Before Use 3 14...
Page 68: ...Section 4 Operation 4 16...
Page 118: ...Section 5 Remote Control 5 50...
Page 136: ...Appendix App ii...
Page 142: ...Appendix A Performance Test Result Recording List A 6...
Page 144: ...Appendix B Error Code B 2...