Section 5
Remote Control
5-6
MFRame:POWer:STATe
Function
Sets optical output state.
Program message
MFRame:POWer:STATe
< sw >, < ch - list >
MFRame:POWer:STATe? < ch - list >
Response message
< sw > {, < sw >}
∗
Parameter
< sw >
:= {ON, OFF, 1, 0}
< ch - list >
:= (@< channel >{,< channel >}
∗
)
< channel >
:= {n|1
≤
n
≤
9}
Explanation
This command turns the optical output on and off.
The response messages are outputted in the order of the channel list.
Example
MFR:POW:STAT 1, (@2, 5, 8)
Turns on the light sources of 2, 5 and 8 channels.
MFRame:POWer:FREQuency
Function
Sets modulation frequency.
Program message
MFRame:POWer:FREQuency CW | < freq > [< unit >], < ch - list >
MFRame:POWer:FREQuency? < ch - list >
Response message
< freq > {, < freq >}
∗
Parameter
< freq >
:= {0, 270, 1000, 2000} (Unit: Hz)
< unit >
:= {HZ, KHZ}
< ch - list >
:= (@< channel >{,< channel >}
∗
)
< channel >
:= {n|1
≤
n
≤
9}
Explanation
Set the modulation light frequency to either CW or the modulation frequency
specified in <freq>. If the units in <freq> are omitted, Hz is used as the unit; or if
the unit is defined in <unit>, the defined unit is used. The CW is used when 0 Hz
is defined.
The response messages are outputted in Hz unit in the order of the channel list.
Summary of Contents for MT9812B
Page 11: ...For Safety xi Laser Radiation Markings D E C A B G F 1...
Page 20: ...xx...
Page 24: ...IV...
Page 38: ......
Page 52: ...Section 3 Before Use 3 14...
Page 68: ...Section 4 Operation 4 16...
Page 118: ...Section 5 Remote Control 5 50...
Page 136: ...Appendix App ii...
Page 142: ...Appendix A Performance Test Result Recording List A 6...
Page 144: ...Appendix B Error Code B 2...