5-37
Remote Control
5
5.9
IEEE488.2 Common Commands
The following commands are used only for GPIB control.
∗
CLS
Function
Clears the status.
Program message
∗
CLS
Explanation
The
∗
CLS common command clears all event registers and queues except an out-
put queue and its MAV summary messages, thus clearing the corresponding sum-
mary messages.
Values set in enable registers are not changed by the
∗
CLS command.
∗
ESE
Function
Enables the standard event status.
Program message
∗
ESE < mask >
∗
ESE?
Response message
< mask >
Parameter
< mask >
:= {Logical sum of 2
n
|0
≤
n
≤
7}
Explanation
The sum total of values (2
0
= 1, 2
1
= 2, 2
2
= 4, 2
3
= 8, 2
4
= 16, 2
5
= 32, 2
6
= 64, and/
or 2
7
= 128) corresponding to the standard event status enable register bits 1, 2, 3,
4, 5, 6, and/or 7 that are to be enabled becomes the parameter. The value of the bit
to be disabled is 0.
Register contents
Set value
Bit 7
Power On
128
Bit 6
Not used
64
Bit 5
Command Error
32
Bit 4
Execution Error
16
Bit 3
Device Dependent Error
8
Bit 2
Query Error
4
Bit 1
Not used
2
Bit 0
Operation Complete
1
5.9
IEEE488.2 Common Commands
Summary of Contents for MT9812B
Page 11: ...For Safety xi Laser Radiation Markings D E C A B G F 1...
Page 20: ...xx...
Page 24: ...IV...
Page 38: ......
Page 52: ...Section 3 Before Use 3 14...
Page 68: ...Section 4 Operation 4 16...
Page 118: ...Section 5 Remote Control 5 50...
Page 136: ...Appendix App ii...
Page 142: ...Appendix A Performance Test Result Recording List A 6...
Page 144: ...Appendix B Error Code B 2...