SENSe<n>:AVERage:COUNt
Function
Sets the number of averagings.
Program message
SENSe<n>:AVERage:COUNt < count >
SENSe<n>:AVERage:COUNt?
Response message
< count >
Parameter
< count >
:= {1, 2, 5, 10, 20, 50, 100, 200, 500, 1000}
Explanation
This command sets the number of averagings for a averaging processing. When
the measured light contains noise components, the S/N ratio can be improved by
the averaging processing.
When “1” is set in the <count>, the averaging processing is stopped.
SENSe<n>:BANDwidth:AUTO
Function
Sets the light-receive bandwidth.
Program message
SENSe<n>:BANDwidth:AUTO < sw >
SENSe<n>:BANDwidth:AUTO?
Response message
< status >
Parameter
< sw >
:= {ON, OFF, 1, 0}
< status >
:= {1, 0}
Explanation
This command switches whether or not the automatic setting of light-receive-
circuit bandwidth is enabled.
The MT9812B switches the measuring level range between “AUTO range” and
“MANUAL range.”
The measuring range of the optical sensor unit is provided with two modes; one is
“AUTO range” which automatically changes the measuring range in accordance
with the input optical level, and the other is “MANUAL range” which measures
in an arbitrary specified fixed range.
“AUTO range” is set when the parameter is “ON” or “1”, and “MANUAL range”
is set when it is “OFF” or “0.”
When using in “MANUAL range” mode, specify the bandwidth using “SENSe
<n>:BANDwidth.”
Summary of Contents for MT9812B
Page 11: ...For Safety xi Laser Radiation Markings D E C A B G F 1...
Page 20: ...xx...
Page 24: ...IV...
Page 38: ......
Page 52: ...Section 3 Before Use 3 14...
Page 68: ...Section 4 Operation 4 16...
Page 118: ...Section 5 Remote Control 5 50...
Page 136: ...Appendix App ii...
Page 142: ...Appendix A Performance Test Result Recording List A 6...
Page 144: ...Appendix B Error Code B 2...