5-13
Remote Control
5
5.3
Device Message
SENSe<n>:POWer:RANGe[:UPPer]
Function
Sets manual range.
Program message
SENSe<n>:POWer:RANGe[:UPPer] < level > [DBM]
SENSe<n>:POWer:RANGe[:UPPer]?
Response message
< level >
Parameter
< level >
:= {30, 20, 10, 0, –10, –20, –30, –40, –50, –60, –70, –80, –90, –100, –110}
Explanation
This command executes measurements in the fixed <level> measuring range
when the “MANUAL range” mode is set in the “SENSe<n>:POWer:RANGe
:AUTO” setting. The units are “DBM” only or the units can be omitted.
POINT
• In the case of “AUTO range” mode, a certain time lag
is unavoidable for switching the range. It is therefore
recommended that the measurements be made in the
“MANUAL range” mode when the optical level to be
measured is roughly known or the variation of its
level is not significant (2 to 3 dB).
• When you switch from “AUTO range” mode to
“MANUAL range” mode, the measuring range be-
comes the last range of the “AUTO range”, and not
that of “MANUAL range” which was set previously.
• The <level> depends on the “optical power measuring
range” of the unit currently in use and the bandwidth
that is set in “SENSe<n>:BANDwidth.” It may not be
possible to set some <level> values in a unit. Refer to
paragraph “2.2 Specifications” and descriptions of
bandwidth setting for the range of each unit currently
in use.
Summary of Contents for MT9812B
Page 11: ...For Safety xi Laser Radiation Markings D E C A B G F 1...
Page 20: ...xx...
Page 24: ...IV...
Page 38: ......
Page 52: ...Section 3 Before Use 3 14...
Page 68: ...Section 4 Operation 4 16...
Page 118: ...Section 5 Remote Control 5 50...
Page 136: ...Appendix App ii...
Page 142: ...Appendix A Performance Test Result Recording List A 6...
Page 144: ...Appendix B Error Code B 2...