Biometra TAdvanced
Protocol Wizard
83
10.2
Enter a new polymerase or edit a polymerase
To enter program templates for new polymerases or to edit preset program templates
for polymerases press the corresponding button:
New polymerase
Edit polymerase
In the next screen (see Fig. 64) touch the corresponding input boxes to enter a name for
the polymerase and set parameters for all requested program steps. With two excep-
tions for most program steps the temperature and the hold time must be defined:
The temperature for the annealing step depends on the calculated primer melting
temperature (T
m
value) and has to be set individually for each primer pair (see sec-
tion "Primer Annealing Temperature Calculator" p. 87).
The hold time for the elongation step becomes calculated by the expected product
length.
The corresponding input boxes for the annealing temperature and elongation time are
shown in grey and cannot be activated.
Fig. 64 Edit polymerase screen
After creating the program template for a new polymerase or editing the program tem-
plate of a polymerase press the
done
button.
Done
In the next screen (see Fig. 65) touch the corresponding button to select a storage place
for the program template (polymerase). If the number of polymerases exceeds the
Summary of Contents for Biometra TAdvanced 384
Page 8: ...Figures Biometra TAdvanced 8...
Page 46: ...Operation Biometra TAdvanced 46 Fig 21 Multiblock selection query 2...
Page 73: ...Biometra TAdvanced Starting coping and deleting programs 73 Fig 49 Confirmation prompt...
Page 122: ...Quick reference guide Biometra TAdvanced 122 Spreadsheet view Graphical view...
Page 138: ...Declaration of conformity Biometra TAdvanced 138 19 Declaration of conformity...