Quick reference guide
Biometra TAdvanced
120
To switch the heated lid preheating on or off, press the corresponding button.
Select a program step and press
Edit
. The screen for editing program steps opens:
Use the cursor keys above the keyboard to switch between program steps. The num-
ber of the currently activated step is displayed on a blue background in the table at
the top of the screen.
Touch the
Temperature
input field and enter a temperature
°C
.
Touch the
Hold time
input field and enter a time. To program a pause step press
∞
.
To program a loop, touch the
Go To
input field and enter the step number to which
the program should return. Then touch the
Cycles
input field and enter how often
the loop should be repeated.
To reduce or increase the annealing temperature from step to step within a loop,
press
∆
T
and set a negative or positive temperature increment.
To reduce or increase the hold time from step to step within a loop, press
∆
t
and set
a time increment.
To adjust the heating rate, press
∆
R
and enter a value between 0.1 °C and max. You
can accept the heating rate for all steps in the program by activating the
Apply heat-
ing/cooling rate to all steps
checkbox.
To program a gradient, press
Grad
. On the next screen you can choose between
standard
Std.
and linear gradient programming
Lin.
. For standard programming en-
ter a temperature for both the first and last row in the sample block. For linear pro-
gramming enter the primer annealing temperature and set a temperature increment.
The gradient curve is displayed. Press
to confirm your entries.
Programming of linear gradient at TAdvanced Mono
at TAdvanced Twin
Press
on the screen to accept the settings. The software leaves the screen and
opens the spreadsheet or graphical program display.
Alternatively, individual program parameters for a step can be edited. In the spread-
sheet view press directly on the cell in the program table which you want to edit. In
For gradient-capable de-
vices only
Summary of Contents for Biometra TAdvanced 384
Page 8: ...Figures Biometra TAdvanced 8...
Page 46: ...Operation Biometra TAdvanced 46 Fig 21 Multiblock selection query 2...
Page 73: ...Biometra TAdvanced Starting coping and deleting programs 73 Fig 49 Confirmation prompt...
Page 122: ...Quick reference guide Biometra TAdvanced 122 Spreadsheet view Graphical view...
Page 138: ...Declaration of conformity Biometra TAdvanced 138 19 Declaration of conformity...