Biometra TAdvanced
Control elements
27
NOTICE
The strength of the lid contact pressure has been designed for a fully populated block. If
only a few samples are to be used in the block, place an additional two (empty) tubes of
the same height in the four corner positions of the block (see Fig. 6). Otherwise the
sample tube may be damaged from excess contact pressure.
Never attempt to close the heated lid with force! This may damage the device.
Fig. 6
Placement of sample tubes in different block systems
5.4.2
Opening the lid
NOTICE
Never open the lid under pressure!
Always follow the sequence below when opening the lid. If the lid is opened under pres-
sure, the locking mechanism may be damaged.
Reduce the contact pressure of the lid completely by turning the lid wheel anticlock-
wise until there is no resistance.
Open the lid by pressing the button on the front of the lid.
The lid opens and locks in a defined end position.