Tools
Biometra TAdvanced
112
96 well sample block (Biometra TAdvanced Mono)
48 well sample block (Bioemtra TAdvanced Twin)
Fig. 95 Sample block loaded with plasticware for the extended self-test
When the plasticware has been properly inserted and the heated lid is closed, the ex-
tended self-test can be started. To start the extended self-test press:
Start Self Test
Fig. 96 Extended self-test screen
The software provides a progress overview during the extended self-test. When a test
has finished, the result
Passed
or
Failed
is displayed. For the currently running test, the
status
Running
is shown and all other tests for which no result is shown are not finished
yet.
Note
: The extended self-test will take approx. 30 minutes. The device is blocked while
the test is running and no other program can be started. To stop the extended self-test,
press the
Cancel Self Test
button.