Creating, editing and saving programs
Biometra TAdvanced
58
Fig. 31 Linear Gradient Tool screen (Biometra TAdvanced Mono)
To create a gradient using the standard gradient programming screen, enter a tempera-
ture for the first column and the last column in the sample block (see Fig. 32). The ef-
fective temperatures for all the other columns are automatically calculated and displayed
as a bar chart.
Fig. 32 Standard gradient programming screen (Biometra TAdvanced Mono)
For Biometra TAdvanced twinblock models
Note
: To be able to use the gradient function, the sample block needs to be gradient en-
abled. For Biometra TAdvanced Twin 48 G this is the left block.
The gradient function is most often used to optimize new primer pairs. A temperature
gradient is established across the sample block in the annealing step so that there are
different temperatures from line to line: