Biometra TAdvanced
Creating, editing and saving programs
51
Fig. 26 Program template screen
7.3
Assigning program names
Each template has a specific name which is superimposed in the programming screen.
To enter a name, touch the
Name
input field and enter a new name using the Biometra
TAdvanced keyboard.
Note
: If a character is pressed, the pre-displayed name will be deleted and a new pro-
gram name can be set. If the program name should be modified, please first press an-
other button like delete or space.
7.4
Setting the heated lid temperature
To set the heated lid temperature, touch the
Lid
input field. A small window opens in
the middle of the screen with a keyboard (see Fig. 27). Enter a value between 30 and
110 using the Biometra TAdvanced keyboard.
Note
: The default setting for the heated lid temperature is 99 °C. 99 °C is sufficient for
typical PCR programs.
Use the corresponding buttons to switch the heated lid on or off.