Creating, editing and saving programs
Biometra TAdvanced
54
Fig. 28 Program step editing screen
Note
: To program a step at least the temperature and the hold time must be defined.
3.
To set the temperature, touch the corresponding input field and enter a value be-
tween 3.0°C and 99.0°C. Use the
tab
button or activate the next input field and enter
a hold time in
h:mm:ss
.
The Biometra TAdvanced software allows several program steps to be programmed
subsequently (Multi Step Programming).
4.
Use the cursor keys to navigate between program steps.
Next step
Previous step
The currently activated step is highlighted in blue in the table above the input fields
(see Fig. 28 on page 52).
Note
: By navigating between program steps it is not necessary to confirm the settings
for each single parameter or step and to leave the screen for editing program steps. In-
stead, you can fill in the parameters for all steps subsequently and confirm your entries
after all steps are programmed.
7.6.1.1
Programming loops
A typical PCR program consists of repetitive steps for denaturation, annealing and ex-
tension. For the repetition of steps the Biometra TAdvanced software allows the pro-
gramming of loops. For a loop two parameters have to be set:
The step number to which the program should return (
Go To
)
The number of repetitions (
Cycles
)
1.
To program a loop, select the last step of the loop and enter the step number to
which the program should return in the
Go To
input field (see Fig. 28).
Summary of Contents for Biometra TAdvanced 384
Page 8: ...Figures Biometra TAdvanced 8...
Page 46: ...Operation Biometra TAdvanced 46 Fig 21 Multiblock selection query 2...
Page 73: ...Biometra TAdvanced Starting coping and deleting programs 73 Fig 49 Confirmation prompt...
Page 122: ...Quick reference guide Biometra TAdvanced 122 Spreadsheet view Graphical view...
Page 138: ...Declaration of conformity Biometra TAdvanced 138 19 Declaration of conformity...