Biometra TAdvanced
Safety instructions
13
2.6
Safety instructions – operation
2.6.1
General
The operator of the Biometra TAdvanced must make sure before each commissioning
that the condition of the device including the safety equipment is sound.
The following has to be observed:
Free access to the power switch on the back of the enclosure has to be ensured dur-
ing operation.
The ventilation fittings at the rear of the device must be unobstructed and opera-
tional. Covered ventilation grilles or slits etc. may cause the device to break down or
may cause damage to it.
The use of oil between the samples and the sample block is not necessary to achieve
an improved heat exchange. However, if you still want to use oil, you should use
mineral oil. Do not use silicone oil.
Be careful to avoid crushing or pinching injuries when closing the device.
The Biometra TAdvanced may cause burns. The following has to be observed:
The thermal block, the samples and the heated lid reach high temperatures. There is
a risk of burns during contact.
Ensure that the lid is securely closed before starting the program! Do not touch the
heated lid!
The rapid heating of the thermal block can cause liquids to boil explosively. Always
wear safety goggles during operation!
Do not touch hot sample tubes or plates and do not open them or boiling liquid may
escape!
Only use plates and tubes that are suitable for high temperatures (up to 100°C), fit
well into the thermal block (no shaking) and whose lids seal tightly!
2.6.2
Safety instructions: protection against explosion and fire
The Biometra TAdvanced must not be operated in an explosive environment.
The Biometra TAdvanced must not be operated with flammable, explosive or volatile
substances.
2.6.3
Safety instructions: electrical equipment
Work on the electrical components of the Biometra TAdvanced may only be performed
by a qualified electrical technician in accordance with the applicable electrical engineer-
ing rules. Life-threatening electrical voltages may occur in the interior of the Biometra
TAdvanced!
The following has to be observed:
Any work on the interior of the device may only be carried out by the customer ser-
vice of Biometra GmbH and specially authorized technicians.
Summary of Contents for Biometra TAdvanced 384
Page 8: ...Figures Biometra TAdvanced 8...
Page 46: ...Operation Biometra TAdvanced 46 Fig 21 Multiblock selection query 2...
Page 73: ...Biometra TAdvanced Starting coping and deleting programs 73 Fig 49 Confirmation prompt...
Page 122: ...Quick reference guide Biometra TAdvanced 122 Spreadsheet view Graphical view...
Page 138: ...Declaration of conformity Biometra TAdvanced 138 19 Declaration of conformity...