Creating, editing and saving programs
Biometra TAdvanced
52
Fig. 27 Heated lid temperature setting screen
NOTICE
The heated lid automatically switches off at a temperature difference of >75 °C between
the block and the heated lid. At these low block temperatures, sample condensation on
the tube lid is no longer to be expected. In addition, it supports a longer service life of
the Peltier elements in the block, since no unnecessary cooling against the heating cover
is required.
7.5
Preheating the heated lid
The heated lid can be set to preheating ON or OFF using the corresponding buttons:
Preheat heated lid Off
Preheat heated lid On
If the preheating mode is activated, the heated lid is first heated up to the programmed
temperature while the sample block is held constant at 25°C. After the heated lid
reaches the target temperature, a 40 second equilibration phase follows. Then the sam-
ple block starts to heat up from 25°C to the first programmed target temperature.
Note:
For most applications it is recommendable to set the preheating mode to "ON".
7.6
Editing steps
When editing program steps, the following parameters can be entered in the Biometra
TAdvanced software: